BECKMAN COULTER INC

UNICEL DXI 800 · Hematology

510(k): 2 Recall: 10 UDI: 0 MAUDE: 10 Class.: 56

510(k) Clearances (2)

2003-01-28

Substantially Equivalent

Recalls (10)

2025-10-01

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

2018-12-05

When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely …

2011-01-26

The recall was initiated due to Beckman Coulter receiving a report of results that were associated with incorrect patient identifications when two sample racks advanced into the sample presentation area in one push. The first rack of the two was not scanned. This is referred to as a "pushed pair." If a pushed pair event occurs, results may be associated with the incorrect patient identification.

2010-10-22

This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.

2010-10-22

This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.

2010-10-22

This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.

2010-08-13

This recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated as a result of missing or incorrect assay reagents. In these cases both false positive and false negative results are possible, depending upon the assay configuration.

2009-10-28

This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample processing, the system will post a yellow caution event and continue to operate with the dispense plate in an elevated position until the instrument

2008-10-09

The recall was initiated after Beckman Coulter confirmed reports of a new premature failure of the upper aspirate peristaltic pump in the systems listed above. Premature failure of the pump may negatively affect precision. The failure is limited to those UniCel DXI 800, DXI 600, and DxC 880i systems which have a pump identified with one of the new pump manufacturing numbers (Mfg. Nos.) listed ab

MAUDE Adverse Events (10)

Malfunction
20080205
Malfunction
20080205
Malfunction
20080203
Malfunction
20080131
Malfunction
20080126
Malfunction
20080125
Malfunction
20080124
Other
20080124
Malfunction
20080117
Other
20080114

FDA Classification (56)

Assay, Heparin

Hematology

Adamts13 Activity Test System

Hematology

Stain, Papanicolau

Hematology

Tablet, Resazurin

Hematology

Solution, Silver Carbonate

Hematology

Ethyl Eosin

Hematology

Lymphocyte Separation Medium

Hematology

Microscope, Automated, Image Analysis, Operator Intervention

Hematology

Test, Sickle Cell

Hematology

Nigrosin

Hematology

Control, Platelet

Hematology

Oxyhemoglobin

Hematology

Plasma, Coagulation Control

Hematology

Indigocarmine

Hematology

Antithrombin Iii Quantitation

Hematology

Dna-Probe, B And T Lymphocyte

Hematology

Stains, Chemical Solution

Hematology

Neutral Red

Hematology

Reagent, Schiff

Hematology

Reagent, Russel Viper Venom

Hematology

Romanowsky Stains

Hematology

Centrifuge, Microsedimentation

Hematology

Giemsa Stain

Hematology

Adenine Nucleotide Quantitation

Hematology

Stains, Microbiologic

Hematology

Separator, Automated, Blood Cell, Diagnostic

Hematology

Centrifuge, Cell-Washing, Automated For Immuno-Hematology

Hematology

Analyzer, Chromosome, Automated

Hematology

Immunohistochemistry Reagents And Kits

Hematology

Products, Red-Cell Lysing Products

Hematology

Carbol Fuchsin

Hematology

Counter, Differential Cell

Hematology

Control Material, Blood Circulating Epithelial Cancer Cell

Hematology

Hemoglobinometer

Hematology

Crystal Violet For Histology

Hematology

Warmer, Blood, Non-Electromagnetic Radiation

Hematology

System, Automated Platelet Aggregation

Hematology

Whole Blood Hemoglobin Determination

Hematology

Jenner Stain

Hematology

Red-Cell Count By Photometry

Hematology

Stains, Heinz Body

Hematology

Fluid, Diluting, Red-Cell

Hematology

Reagent, Thromboplastin And Control

Hematology

Hematoxylin, Chrome Alum

Hematology

Reagent & Control, Partial Thromboplastin Time

Hematology

Hemoglobin For Donor Testing

Hematology

Carmine

Hematology

Bothrops Atrox Reagent

Hematology

Vitamin K Epoxide Reductase Complex Subunit One (Vkorc1) Genotyping System

Hematology

Auramine O

Hematology

Assay, Glycosylated Hemoglobin

Hematology

Mucicarmine

Hematology

Control, Hemoglobin

Hematology

Red-Cell Indices, Measured

Hematology

Blood Establishment Computer Software And Accessories

Hematology

Van Gieson'S Picro-Fuchsin

Hematology