CAPINTEC INC
CAPTUS 4000E · Gastroenterology and Urology
Recalls (6)
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
Complaints of unexpected detachment of the collimator have been reported.
Complaints of unexpected detachment of the collimator have been reported.
Complaints of unexpected detachment of the collimator have been reported.
Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.
UDI Registrations (3)
The Captus 4000e is a PC based Thyroid Uptake System, on a ergonomically designed mobile stand. It includes a 1024 channel MCA, automated Quality Assurance, Thyroid Uptake Module, and Bioassay module. This model includes a well counter with 2 inch of lead shielding.
The Captus 4000e is a PC based Thyroid Uptake System, on a ergonomically designed mobile stand. It includes a 1024 channel MCA, automated Quality Assurance, Thyroid Uptake Module, and Bioassay module.
The Captus 4000e is a PC based Thyroid Uptake System, on a ergonomically designed mobile stand. It includes a 1024 channel MCA, automated Quality Assurance, Thyroid Uptake Module, Bioassay module, Wipe Test Module, and Lab Tests.