BECKMAN COULTER INC
REMISOL ADVANCE · General and Plastic Surgery
510(k): 0
Recall: 2
UDI: 3
MAUDE: 1
Class.: 0
Related Cybersecurity Alerts
Recalls (2)
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and …
There is a potential that the data management system may add additional cells to the patient request which could lead to erroneous patient result.
UDI Registrations (3)
REMISOL Advance
REMISOL Advance, version 2.3
REMISOL Advance
REMISOL Advance, version 2.2
REMISOL Advance
REMISOL Advance
MAUDE Adverse Events (1)
Malfunction