HOSPIRA INC
LIFECARE PCA · Anesthesiology
510(k) Clearances (3)
Substantially Equivalent
Substantially Equivalent
Substantially Equivalent
Recalls (10)
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.
Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.
The infusion pump has a clear plastic door that, due to its design, allows for excessive flexing. This in turn permits unauthorized access to the vial of medication.
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Incorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.
MAUDE Adverse Events (10)
FDA Classification (59)
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology
Anesthesiology