SIEMENS HEALTHCARE DIAGNOSTICS INC
RAPIDPOINT 500 · Cardiovascular
510(k) Clearances (1)
Substantially Equivalent
Recalls (10)
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufact…
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as ca…
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as ca…
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as ca…
RAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as ca…
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a de…
Potential Patient Demographic Error with Blank Patient ID Field
Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
RAPIDPoint 500 Measurement Cartridges may have an error code one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-)
RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
MAUDE Adverse Events (10)
FDA Classification (63)
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular
Cardiovascular