DYNEX TECHNOLOGIES INC
AGILITY · Hematology
Recalls (2)
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.
Control samples aspirated from wrong SmartKit on the Agility. Agility software was updated to v1.4.7 to resolve the issue. This leads to a risk that a control from another assay's SmartKit will be used instead of the correct control, which may lead to delayed patient results.
UDI Registrations (1)
Automated ELISA Processing System with 12 Incubators
FDA Classification (56)